top of page

MITUTOYO 544-538 Laser Scan Micrometer LSM-506S Measuring Unit
Order No.
544-538
Product Details
Technical Data:
Series: 544
Model: LSM-506S
Applicable laser standards: IEC, FDA
User's manual: English version
Measuring range: 0.040 to 2.36 in / 1 to 60 mm
Resolution: 0.000002 to 0.005 in / 0.05 to 100 um selectable
Repeatability: +/-0.36 um
Accuracy at 20 degC: Whole range +/-3 um; narrow range +/-(1.5 + 0.5 D) um
Positional error: +/-4 um
Measuring area, optical axis x scanning direction: 20 x 60 mm
Scanning rate: 3200 scans/s
Laser wavelength: 650 nm visible
Laser scanning speed: 452 m/s
Operating temperature: 0 to 40 degC
Operating humidity: RH 35 to 85%, non-condensing
Protection rating: IP64
Mass: Emission unit 1.1 kg; reception unit 0.6 kg; base 0.5 kg; signal cable 5 m 0.5 kg
Content:
Signal cable 5 m: 02AGN770A
Application:
Non-contact laser scanning micrometer measuring unit for high-speed outside-diameter and dimensional measurement.
Dimensions:
8 x M3; 80; 5.5; 250; 105; 84.5; 54.5; 42; 170; 80; 50; 55; 42; 8; 48; 259; 202; 57; 6; 18.5; 50; 22; 10.5; 15.5; 9.5; 14.5; 90; 57; 41; 12; 110; 20; 10; 30; 355; 46; 53; 68; 95.5; 65.5; 150; 57.5; 130 mm; max. emission/reception separation 700 mm; relay cable to reception 1.4 m; extendable dimension up to 270 mm.
Remark:
Repeatability is specified as +/-2 sigma under the catalogue test condition.
D in narrow-range accuracy is the difference in diameter between master gage and workpiece in mm.
Positional error is outside-diameter error caused by workpiece-position variation in the optical-axis or scanning direction.
Protection rating applies to the interior; water or dust on the measuring-unit window may cause malfunction.
Optional:
544-047: LSM-5200 panel-mount display
542-072A: EH-102D display, English mm/in
02AGN780A/B/C/D: Extension signal cables, 5/10/15/20 m
02AGC150A/B/C: Extension relay cables, 1/3/5 m
02AGD140 calibration gage set; 02AGD520 adjustable workstage; 02AGD250 air blower
bottom of page